HybridSPE: A novel technique to reduce phospholipid-based matrix effect in LC-ESI-MS Bioanalysis


1 School of Pharmacy and Technology Management, NMIMS University, Mumbai, India
2 Department of Bioanalytical, AshcoNeulab Ltd, Mumbai, India
3 Faculty of Pharmacy, Integral University, Lucknow, India

Date of Submission29-Aug-2011
Date of Decision19-Dec-2011
Date of Acceptance05-Jan-2012
Date of Web Publication07-Nov-2012

Correspondence Address:
Md Akhlaquer Rahman
Faculty of Pharmacy, Integral University, Lucknow
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0975-7406.103234

Rights and Permissions
   Abstract 

When complex biological materials are analyzed without an adequate sample preparation technique, MS signal and response undergo significant alteration and result in poor quantification and assay. This problem generally takes place due to the presence of several endogenous materials component in samples. One of the major causes of ion suppression in bioanalysis is the presence of phospholipids during LC-MS analysis. The phospholipid-based matrix effect was investigated with a commercially available electro spray ionization (ESI) source coupled with a triple quadrupole mass spectrometer. HybridSPE dramatically reduced the levels of residual phospholipids in biological samples, leading to significant reduction in matrix effects. This new procedure that combines the simplicity of precipitation with the selectivity of SPE allows obtaining much cleaner extracts than with conventional procedures. HybridSPE-precipitation procedure provides significant improvement in bioanalysis and a practical and fast way to ensure the avoidance of phospholipids-based matrix effects. The present review outlines the HybridSPE technique to minimize phospholipids-based matrix effects on LC-ESI-MS bioanalysis.

Keywords: Bioanalytical, electro spray ionization, HybridSPE, matrix effects, phospholipids


How to cite this article:
Ahmad S, Kalra H, Gupta A, Raut B, Hussain A, Rahman MA. HybridSPE: A novel technique to reduce phospholipid-based matrix effect in LC-ESI-MS Bioanalysis. J Pharm Bioall Sci 2012;4:267-75

How to cite this URL:
Ahmad S, Kalra H, Gupta A, Raut B, Hussain A, Rahman MA. HybridSPE: A novel technique to reduce phospholipid-based matrix effect in LC-ESI-MS Bioanalysis. J Pharm Bioall Sci [serial online] 2012 [cited 2020 Mar 31];4:267-75. Available from: http://www.jpbsonline.org/text.asp?2012/4/4/267/103234


   Introduction Top


High-performance liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) due to its high degree of sensitivity and selectivity is widely used in bioanalysis of small molecules. When complex materials are analyzed without an adequate sample preparation technique, MS signal and response undergoes significant alteration and result in poor quantification and assay. This problem generally takes place due to the presence of several endogenous materials component in samples. [1],[2],[3],[4],[5],[6],[7] Matrix effect is defined as the effect of co-eluting residual matrix components on the ionization of the target analyte. When a rather long isocratic or gradient chromatographic program is used in the quantitative assay, matrix effect may not be present at the retention time for an analyte. However, in the case of high-throughput LC-MS/MS analysis, matrix effect is one of the major issues to be addressed in method development and validation, especially when analyte is not well separated from the LC-front. One problem brought by matrix suppression effect is reduced sensitivity when analyte signal is suppressed. Detailed studies on matrix effects revealed that the ion suppression or enhancement are frequently accompanied by significant deterioration of the precision of the analytical method as demonstrated by Matuszewski et al.[8] The authors studied the precision (%R.S.D.) upon repetitive injection of post-extraction spiked plasma samples as a function of the analyte concentration for a single lot and for five different lots of plasma. While for the single plasma lot the precision is acceptable, it may not be when different plasma lots are taken into account. Generally, matrix effect impacts more on the low end of calibration curve than the mid range or highly end. When discussing matrix effects, it is useful to discriminate between ion suppression (or enhancement) by the matrix on one hand, and different matrix effects exerted by different sample lots on the other hand. The difference in response between a neat solution sample and the post-extraction spiked sample is called the absolute matrix effect, while the difference in response between various lots of post-extraction spiked samples is called the relative matrix effect. If no counteraction is taken, an absolute matrix effect will primarily affect the accuracy of the method, while a relative matrix effect will primarily affect the precision of the method. [2] One of the best tools for observing matrix effects is the post column infusion technique.


   Matrix Effect Top


Ion-suppression, sometimes referred to as matrix effect, is a common problem in atmospheric pressure ionization (API) mass spectrometer. Phospholipids, and in particular glycerophosphocholines and lysophosphatidylcholine [Figure 1] represent the major class of endogenous compounds causing significant matrix effects. [2],[3],[4],[5],[6],[7],[9] Phospholipids are important class of biological compounds containing one or more phosphate groups.

For their inherent characteristics these polar lipids are of great importance for the structure and function of cell membranes and are the most abundant of membrane lipids. Although not stored in large amounts in the system, [10] they are particularly abundant in plasma samples. Their molecular structure exhibit two major functional group regions: a polar head group substituent, which includes an ionizable organic fatty acid ester group which impart considerable hydrophobicity to the molecule. In particular, the highly ionic nature of phospholipids makes them responsible for influencing the ionization in electrospray MS sources, [11],[12] and desolvation of the LC effluent droplets in electrospray MS analysis. [13]
Figure 1: Structural formula of phospholipids and their fragment contributing major matrix effect problem

Click here to view


Therefore, the removal of phospholipids represents phosphate moiety as well as other polar groups of various types, and one or two long chain being an extremely important step in the sample preparation process. [14] Several techniques have been applied toward the removal of phospholipids in order to obtain cleaner sample extracts. Most of them are based on solid-phase extraction procedures using strong cation exchange sorbents, [12] automated on-line SPE. [15] Matrix effect is a common problem in atmospheric pressure ionization (API) mass spectrometry. [16] There is difference in opinion as to amount of ion-suppression that is acceptable for an analytical method. In some laboratories, ion-suppression in an analytical method is not acceptable, but in other cases, it is tolerated if there is no significant effect on the validity of the analytical data with appropriate QCs. The phenomenon of ion-suppression results in sporadic reduction of signal intensity. Consequently, the lower limit of quantitation (LLOQ) for highly sensitive bioanalytical methods could be difficult to achieve. The problem can further be complicated by differential ion-suppression due to inter subject variability, and the use of blank bio-matrix with varying lot numbers (i.e., control blood obtained from different patients/subjects) in preparing the calibrators. Differences in ion-suppression between the analyte and structurally different internal standards may also be problematic. This issue can be mostly mitigated by the use of stable isotopically labeled (SIL) internal standards. The extent of ion-suppression is dependent on the sample preparation method and chromatographic separation. The use of protein precipitation (PPT) is most likely to cause an ion-suppression in ESI due to the lack of selectivity (co-elution of endogenous compounds such as lipids, phospholipids, fatty acids, etc., that affect the ESI droplet de-salvation process; however, in comparison, the extracts obtained from solid phase extraction (SPE) and liquid-liquid extraction (LLE) are relatively cleaner. [9] Further studies on the molecular identities of co-eluting endogenous compounds are required to clearly delineate their contributions to ion-suppression. Ion-suppression can lead to random variations in the internal standard signal, standard curve slope and intercept, LLOQ, and overall reproducibility of an assay. Lastly, with an exception, the use of APCI instead of ESI can eliminate or minimize ion-suppression. However, the analyst should perform preliminary experiments on assay sensitivity as well as thermal stability of the test article during APCI optimization to ascertain its structural integrity during analysis. [17]


   How to Diagnose that it is Matrix Effect or Poor Recovery Top


Recovery is determined by comparing the peak area of extracted sample with those neat solutions spiked into the blank matrix. Because both samples have the matrix ingredients present, the matrix can be considered same for the extracted sample. Any difference in response is due to the extraction recovery. The extraction efficiency (or recovery) was determined by measuring an extracted sample against a (post-extraction) spiked sample:

Where, A is the response for the extracted samples and B the response for post- extracted spiked samples.

Now the matrix effect is determined by comparing the peak area of the neat solution spiked into blank (post-extracted) with those of the neat solution. Because analyte is not extracted, it should have same response in the post-extraction spiked sample and in neat solution. The matrix effect will be responsible for difference in response. A useful method to assess matrix suppression is post-column infusion of analyte into mass detector. The matrix effect was measured by referring the post-extracted spiked sample to the unextracted sample, using the following equation:

Where, C is the peak area of a neat standard and B is the corresponding peak area for standards spiked into plasma after extraction.


   A Post-Column Infusion Experiment That Can be Used to Detect Matrix Effects in an HPLC-MS-MS Assay Top


Here the extracted blank matrix is injected by an auto sampler into the analytical column. The purpose of post-column infusion with analyte is to raise the background level so that the suppression matrix will show negative peak. This method [Figure 2] has been successfully used to identify matrix suppression peak. [18]
Figure 2: Schematic of post column infusion experiment commonly used for the qualitative assessment of ion suppression

Click here to view



   Problems/Difficulties in Removing Phospholipids from Biological Samples Top


Due to the nature of phospholipids, none of the mainstream sample preparation techniques; protein precipitation technique (PPT), liquid-liquid extraction (LLE), and solid phase extraction (SPE) offer the proper selectivity to distinguish phospholipids from analytes of interest. Their molecular structure exhibit two major functional group regions: a polar head group substituent, which includes an ionizable organic phosphate moiety as well as other polar groups of various types, and one or two long chain fatty acid ester groups [Figure 3], which impart considerable hydrophobicity to the molecule. Phospholipids co-extract with analytes of interest when using traditional sample preparation technique. [19]
Figure 3: Phospholipids co-extracts with analytes of interest when using traditional techniques (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view


Protein precipitation technique

Protein precipitation is the simplest means of sample pretreatment. Due to the selectivity of MS detectors, it is thought that sample pretreatment for LC-MS/MS assays is redundant. Infact, the largely polar and hydrophilic character of some natural products makes it difficult to extract analytes from plasma with solvents, and PPT technique was therefore often used. [20],[21] The results showed the deproteinization by acetonitrile gave a good resolution and high recovery. Based on the above mentioned reasons, many natural products in biological fluids were extracted by this method. [22],[23],[24],[25],[26],[27],[28],[29],[30] It is also worth noting that precipitation of proteins with acids may catalyze the hydrolysis of some conjugates such as glucuronides and sulphates. [31] It removes gross levels of proteins but no phospholipids.

Liquid-liquid extraction

Liquid-liquid extraction is especially suited for lipophilic compounds. The selection of sample pretreatment techniques for each bio-fluid depends on expected analyte concentrations and required detection limits. Acetonitrile protein precipitation provides sufficient pre-concentration and protein removal for quantitative analysis of analytes in bio-fluids. Matrix suppression data indicate that SPE or LLE is required prior to LC-MS bioanalysis. [3] In LLE phospholipids will co-extract with analytes due to hydrophobic tail. [19]

Solid-phase extraction

In the bioanalysis solid-phase extraction is a frequently used technique for sample pre-treatment. Compared with the PPT procedure, the SPE method reduces the serum background greatly [Figure 4]. SPE is chosen for the extraction and purification of analytes due to its high selectivity, speed of extraction, the potential for automation, and the fact that much lower volumes of organic solvents are required than those for liquid-liquid extraction. [32] In some cases, it is necessary to acidify or alkalify the biological fluids containing analytes before transferring samples to the SPE column. [33],[34] SPE can be performed off-line manually; it is known that the efficiency of SPE depends on the type of sorbents, the sample volume and pH, the content of organic modifier. In SPE, nominal phospholipids are removed (in reversed-phase SPE-phospholipids will co-extract with analytes due to hydrophobic tail while in mixed-mode/ion-exchange SPE-phospholipids will co-extract with analytes due to the zwitterionic polar head group). [19]
Figure 4: Different step of SPE (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view



   Novel-Hybrid SPE-Precipitation Technology Top


A pitfall of all the above-discussed methods is the matrix effect since all of these methods, except one, used pure neat analyte solutions to model the SPE characteristics for use with real matrix samples. Since plasma, serum, and urine are complex matrices, containing many compounds that could interfere with analyte retention, care must be taken in applying the models to real samples. To diminish the effect of the matrix, it can be diluted as far as possible before applying it to the sorbent. Also, it is wise to take some margin into account in selecting organic modifier concentrations used in wash and elution steps. The matrix effect is compound dependent and rather unpredictable. Therefore, the recovery of a proposed SPE method should always be verified by using a spiked matrix sample.

HybridSPE-PPT technology is a simple and generic sample prep platform designed for the gross level removal of endogenous protein and phospholipids interferences from biological plasma and serum prior to LC-MS or LC-MS-MS analysis. The important feature and benefits of this technique are mentioned in [Table 1].
Table 1: Important feature and benefits of the HybridSPE-PPT technique

Click here to view


Procedure

Biological plasma or serum is first subjected to protein precipitation in which biological plasma/serum is first added to the 96-well plate followed by acidified acetonitrile (precipitation agent). An upper PTFE frit impedes premature flow of the sample before vacuum application. After a brief mixing/vortexing step to facilitate protein ppt, vacuum is applied to the 96-well plate.

The 96-well version contains a series of low porosity hydrophobic filters/frits. The packed-bed filter/frit assembly acts as a depth filter facilitating the concurrent removal of both phospholipids and precipitated proteins during the extraction process. After the protein-precipitated sample passes through the HybridSPE-PPT device, the resulting eluent is ready for immediate LC-MS or LC-MS-MS analysis [Figure 5].
Figure 5: Different step of HybridSPE-PPT (Courtesy of Sigma-Aldrich; Hybrid SPE-ppt, USA)

Click here to view


Working mechanism

Once the plasma/serum sample is subjected to protein precipitation via the addition of acetonitrile (containing 1% formic acid), it is passed through the HybridSPE-PPT packed bed. The packed bed consists of proprietary zirconia-coated silica particles [Figure 6]. The zirconia sites exhibit Lewis acid (electron acceptor) properties that will interact strongly with Lewis bases (electron donor). Phospholipids structurally consist of a polar head group (zwitterionic phosphonate moiety) and a large hydrophobic tail (two fatty acyl groups that are hydrophobic). The phosphate group inherent with all phospholipids acts as a very strong Lewis base that will interact strongly with zirconia atoms functionalized on the particle surface. Formic acid is a critical modifier used in the precipitation agent to improve the recovery of many analytes of interest (e.g., acidic compounds). It plays a critical role in preventing analyte retention without affecting phospholipids retention/removal.
Figure 6: HybridSPE-ppt 96-well schematic and phospholipids retention mechanism (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view



   Method Development and Recovery Optimization Top


Zr atoms bonded to the HybridSPE silica surface act as Lewis acid because it contains vacant d-orbital and interacts with sample component by Lewis acid-base interaction. Phospholipids are retained/removed from the sample through a highly selectivity Lewis acid-base interaction between the phosphate group (Lewis base), inherent of all phospholipids, and Zr atoms bonded to the HybridSPE silica surface. When developing new HybridSPE-PPT methods, there are three precipitation agents recommended for researchers are citric acid, formic acid and ammonium form a tea modifier that inhibits analyte(s) of interest from co-retaining with phospholipids on the Zr-Si HybridSPE stationary phase, potentially resulting in low recovery.

Formic acid is used as a modifier for most of the acidic, basic, and neutral compound. Because the formate ion is stronger lewis base that can prevent the interaction of most of acid compound to Zerconiasites and H + produced by formic acid ionization neutralized silanol effect, it can improve recovery [Figure 7].
Figure 7: Interaction of formic acid modifier with zerconia-coated stationary phases (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view


Ammonium formate in methanol recommended for low recovery basic and neutral compounds. Methanol combined with ammonium formate is a powerful protein precipitation agent, providing precipitants similar in consistency to acetonitrile protein precipitation. The formate ion (HCOO - ) is a strong enough Lewis base to inhibit most acidic compounds from interacting with Zr sites, improving recovery of acidic compounds. NH4 + is a stronger counter-ion than H + , able to inhibit a wider range of basic compounds from interacting with exposed silanol groups.

Citric acid in acetonitrile recommended for low recovery chelator and acidic chelator compounds (citric acid) is a stronger Lewis base than format, inhibiting most chelator compounds from interacting with Zr ions on the HybridSPE stationary phase [Figure 8]. Note that citrate is not strong enough to disrupt phosphate (phospholipids) binding. [19] The flow chart [Figure 9] is given here under for the convenience of reader.
Figure 8: Interaction of ammonia formate modifier with zerconia-coated stationary phases (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view
Figure 9: HybridSPE-PPT method development (recovery optimization) flow chart

Click here to view


Comparison of sample preparation techniques

HybridSPE-PPT combines the simplicity of protein precipitation with the selectivity of SPE targeting the specific removal of both phospholipids and precipitated proteins from biological plasma. Both phospholipids and protein represent the two largest contributors of ion-suppression in pharmaceutical bioanalysis. If adequately removed prior to LC-MS, analysts can more easily and quickly achieve targeted LLOQ for most applications. Comparison of sample preparation techniques are given in [Table 2].
Table 2: Comparison of sample prep techniques

Click here to view


Comparison of HybridSPE-PPT, protein precipitaiton, and SPE

In this application example, rat plasma samples were spiked with clenbuterol (R(-) and S(+) enantiomers) at the level of 10 ng/mL and extracted using three different procedures: HybridSPE-PPT, Protein PPT, and a 9-step SPE procedure optimized for trace level clenbuterol analysis. The analysis was performed using a chiral stationary phase containing a macrocyclic glycopeptide covalently bound to silica and detection via MS-MS. Comparisons of sample preparation methods were made in terms of the amount of phospholipids in the sample extract and the overall effect on signal response of clenbuterol enantiomers. Absolute recovery was determined against an external standard. Representative chromatograms of each the sample preparation techniques are depicted in [Figure 10]. From the results, it was found that phospholipid contamination levels were highest for protein precipitation resulting in signal suppression levels 70% and 25% for the R(-) and S(+) enantiomers of clenbuterol, respectively. For the SPE procedure, phospholipid contamination was still evident after multiple wash steps, and overall absolute recovery was less than 50%. In contrast, HybridSPE-PPT offered 100% removal of phospholipids resulting in absolute recovery levels of 95%.
Figure 10: Comparative extraction and LC-MS-MS of 10 ng/mL Clenbuterol (R(-) and S(+) enantiomers) in rat plasma (Courtesy of Sigma-Aldrich; HybridSPE-ppt, USA)

Click here to view



   Summary and Conclusion Top


The aforementioned discussion indicated that although plasma protein precipitation can sometimes fulfill the requirements of extraction yield but not adequate to sufficiently remove phospholipids. It removes gross levels of proteins. Moreover, the monitoring of the MRM transitions of phospholipids and the matrix effect evaluation clearly indicated that endogenous phospholipids in both rat and human plasma PPT extracts are a significant tool in order to evaluate phospholipids-based matrix effects in LC-MS/MS analyses leading to ion suppression in bioanalytical determinations. While in Liquid-liquid extraction phospholipids will co-extract with analytes due to hydrophobic tail. A systematic comparison of the different traditional sample preparation techniques to the new sample preparation platform HybridSPE-Precipitation showed to represent a useful and effective tool in providing cleaner extract and in reducing or eliminating phospholipids-based matrix effects.


   Acknowledgements Top


It is great pleasure for us to acknowledge all those who have contributed towards the conception, origin and nurturing of this review article. With a deep sense of gratitude and the respect, we thank to Dr. R.S. Gaud, Dean School of pharmacy and Technology Management, NMIMS University, Mumbai (India), for his constant encouragement and co-operation. We extend thanks to Sigma-Aldrich for giving us permission to use their copyrighted material.

 
   References Top

1.Shou WZ, Naidong W. Post-column infusion study of the 'dosing vehicle effect' in the liquid chromatography/tandem mass spectrometric analysis of discovery pharmacokinetic samples. Rapid Commun Mass Spectrom 2003;17:589-97.  Back to cited text no. 1
[PUBMED]    
2.Ismaiel OA, Halquist MS, Elmamly MY, Shalaby A, Karnes HT. Monitoring phospholipids for assessment of matrix effects in a liquid chromatography-tandem mass spectrometry method for hydrocodone and pseudoephedrine in human plasma. Analyt Technol Biomed Life Sci 2007;859:84-93.  Back to cited text no. 2
[PUBMED]    
3.Dams R, Huestis MA, Lambert WE, Murphy CM. Matrix effect in bio-analysis of illicit drugs with LC-MS/MS: Influence of ionization type, sample preparation, and biofluid. J Am Soc Mass Spectrom 2003;14:1290-4.  Back to cited text no. 3
[PUBMED]    
4.Taylor PJ. Matrix effects: The achilles heel of quantitative high-performance liquid chromatography-electrospray-tandem mass spectrometry. Clin Biochem 2005;38:328-34.  Back to cited text no. 4
[PUBMED]    
5.van De Steene JC, Mortier KA, Lambert WE. Tackling matrix effects during development of a liquid chromatographic-electrospray ionization tandem mass spectrometric analysis of nine basic pharmaceuticals in aqueous environmental samples. J Chromatogr A 2006;1123:71-81.  Back to cited text no. 5
[PUBMED]    
6.Jemal M, Xia YQ. LC-MS Development strategies for quantitative bioanalysis. Curr Drug Metab 2006;7:491-502.  Back to cited text no. 6
[PUBMED]    
7.Souverain S, Rudaz S, Veuthey J. Matrix effect in LC-ESI-MS and LC-APCI-MS with off-line and on-line extraction procedures. J Chromatogr A 2004;1058:61-6.  Back to cited text no. 7
    
8.Matuszewski BK, Constanzer ML, Chavez-Eng CM. Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC−MS/MS. Anal Chem 2003;75:3019-30.  Back to cited text no. 8
[PUBMED]    
9.Little JL, Wempe MF, Buchanan CM. Liquid chromatography-mass spectrometry/mass spectrometry method development for drug metabolism studies: Examining lipid matrix ionization effects in plasma. J Chromatogr B 2006;833:219-30.  Back to cited text no. 9
    
10.Myher JJ, Kuksis A, Pind S. Molecular species of glycerophospholipids and sphingomyelins of human plasma: Comparison to red blood cells. Lipids 1989;24:408-18.  Back to cited text no. 10
    
11.Mei H, Korfmacher M. Using Mass Spectrometry for Drug Metabolism StudiesBoca Raton, FL: CRC Press; 2005. p. 103-50.  Back to cited text no. 11
    
12.Shen JX, Motyka RJ, Roach JP, Hayes RN. Minimization of ion suppression in LC-MS/MS analysis through the application of strong cation exchange solid-phase extraction (SCX-SPE). J Pharm Biomed Anal 2005;37:359-67.  Back to cited text no. 12
    
13.King R, Bonfiglio R, Fernandez-Metzler C, Miller-Stein C, Olah T. Mechanistic investigation of ionization suppression in electrospray ionization. J Am Soc Mass Spectrom 2000;11:942-50.  Back to cited text no. 13
    
14.Chambers E, Wagrowski-Diehl DM, Lu Z, Mazzeo JR. Systematic and comprehensive strategy for reducing matrix effects in LC/MS/ MS analyses. J Chromatogr B 2007;852:22-34.  Back to cited text no. 14
    
15.Marchese A, McHugh C, Kehler J, Bi H. Determination of pranlukast and its metabolites in human plasma by LC/MS/MS with PROSPEKT TM on-line solid-phase extraction. J Mass Spectrom 1998;33:1071-9.  Back to cited text no. 15
    
16.Bonfiglio R, King RC, Olah T, Merkle K. The effects of sample preparation methods on the variability of the electrospray ionization response for model drug compounds. Rapid Commun Mass Spectrom 1999;13:1175-85.  Back to cited text no. 16
    
17.Liu DQ, Pereira T. Interference of a carbamoylglucuronides metabolite in quantitative liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom 2002;16:142-6.  Back to cited text no. 17
    
18.Pucci V, Palma SD, Alferi A, Bonelli F, Monteagudo E. A novel strategy for reducing phospholipids-based matrix effect in LC-ESI-MS bioanalysis by means of HybridSPE. J Pharm Biomed Anal 2009;50:867-71.  Back to cited text no. 18
    
19.Sigmaaldrich.com. USA: Sigma-Aldrich. Available from: http://www.sigmaaldrich.com/Graphics/Supelco/objects/4600/4538.pdf[Last accessed on 2010 Jan 12].  Back to cited text no. 19
    
20.Guo JF, Zhao YM, Zhao L, Zhang WQ, Zhang AJ, Xu B. Simultaneous quantification of CTN986 and its deglycosylation products in rat serum using liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom 2006;20:1701-8.  Back to cited text no. 20
    
21.Xing J, Chen X, Zhong D. Stability of baicalin in biological fluids in vitro. J Pharm Biomed Anal 2005;39:593-600.  Back to cited text no. 21
    
22.Ji HY, Lee HW, Kim HK, Kim HH, Chang SG, Sohn DH, et al. Simultaneous determination of ginsenoside Rb(1) and Rg(1) in human plasma by liquid chromatography-mass spectrometry. J Pharm Biomed Anal 2004;35:207-12.  Back to cited text no. 22
    
23.Soepenberg O, de Bruijn P, Verweij J, Atsumi R, Sparreboom A. Liquid chromatographic assays for DE-310, a novel camptothecinanalog, and two major enzymatic products in human matrices. J Chromatogr B 2004;799:15-22.  Back to cited text no. 23
    
24.Zhao HQ, Qu Y, Wang XY, Lu XY, Zhang XH, Hattori M. Determination of trigonelline by HPLC and study on its pharmacokinetics. Yao XueXueBao 2003;38:279-82.  Back to cited text no. 24
    
25.Kaneko R, Hattori S, Furuta S, Hamajima M, Hirata Y, Watanabe K, et al. Sensitive analysis of aconitine, hypaconitine, mesaconitine and jesaconitine in human body fluids and Aconitum-tubers by LC/ ESI- TOF-MS.J Mass Spectrom 2006;41:810-4.  Back to cited text no. 25
    
26.Guo W, Ahmad A, Khan S, Dahhani F, Wang YF, Ahmad I. Determination by liquid chromatography with fluorescence detection of total 7-ethyl-10-hydroxy-camptothecin (SN-38) in beagle dog plasma after intravenous administration of liposome-based SN-38 (LE-SN38). J Chromatogr B 2003;79:85-92.  Back to cited text no. 26
    
27.Zhai S, Sausville E, Figg WD. A high-performance liquid chromatography method using ultraviolet detection for the quantitation of flavopiridol from human plasma. Biomed Chromatogr 2002;16:379-82.  Back to cited text no. 27
    
28.Wang XL, Xing DM, Wang W, Su H, Tao JL, Du LJ. Pharmacokinetics of Berberine in Rat Thalamus after Intravenous Administration of Coptidis Rhizoma Extract. Am J Chin Med 2005;33:935-43.  Back to cited text no. 28
    
29.Cheng X, Shin YG, Levine BS, Smith AC, Tomaszewski JE, van Breemen RB. Quantitative analysis of betulinic acid in mouse, rat and dog plasma using electrospray liquid chromatography/mass spectrometry. Rapid Commun Mass Spectrom 2003;17:2089-92.  Back to cited text no. 29
    
30.Xu RN, Fan L, Rieser MJ, El-Shourbagy TA. Recent advances in high-throughput quantitative bioanalysis by LC-MS/MS. J Pharm Biomed Anal 2007;44:342-55.  Back to cited text no. 30
    
31.Tan ZR, Ouyang DS, Zhou G, Wang LS, Li Z, Wang D, et al. Sensitive bioassay for the simultaneous determination of pseudoephedrine and cetirizine in human plasma by liquid-chromatography-ion trap spectrometry. J Pharm Biomed Anal 2006;42:207-12.  Back to cited text no. 31
    
32.Murphy CM, Huestis MA. LC-ESI-MS/MS analysis for the quantification of morphine, codeine, morphine-3-β-D-glucuronide, morphine-6-β-D-glucuronide, and codeine-6-β-D-glucuronide in human urine. J Mass Spectrom 2005;40:1412-6.  Back to cited text no. 32
    
33.Samanidou VF, Evaggelopoulou EN, Papadoyannis IN. Simultaneous determination of quinine and chloroquine anti-malarial agents in pharmaceuticals and biological fluids by HPLC and fluorescence detection. J Pharm Biomed Anal 2005;38:21-8.  Back to cited text no. 33
    
34.Xing J, Xie C, Lou H. Recent application of liquid chromatography-mass spectrometry in natural product. J Pharm Biomed Anal 2007;44:368-78.  Back to cited text no. 34
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8], [Figure 9], [Figure 10]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 The use of a restricted access volatile supramolecular solvent for the LC/MS-MS assay of bisphenol A in urine with a significant reduction of phospholipid-based matrix effects
José Ángel Salatti-Dorado,Noelia Caballero-Casero,María Dolores Sicilia,María Loreto Lunar,Soledad Rubio
Analytica Chimica Acta. 2017; 950: 71
[Pubmed] | [DOI]
2 Novel strategy for the facile enrichment of isopentenyl pyrophosphate in rat plasma via Ti4+ -immobilized polydopamine@Fe3 O4 core-shell microspheres
Chen Xu,Yingjia Yu,Li Ling,Xueni Sun,Yan Li,Gengli Duan
Journal of Separation Science. 2017;
[Pubmed] | [DOI]
3 Sample Preparation Strategies for the Effective Quantitation of Hydrophilic Metabolites in Serum by Multi-Targeted HILIC-MS/MS
Elisavet Tsakelidou,Christina Virgiliou,Lemonia Valianou,Helen Gika,Nikolaos Raikos,Georgios Theodoridis
Metabolites. 2017; 7(2): 13
[Pubmed] | [DOI]
4 Solid-phase extraction of the alcohol abuse biomarker phosphatidylethanol using newly synthesized polymeric sorbent materials containing quaternary heterocyclic groups
Mariana Duarte,Kishore Kumar Jagadeesan,Johan Billing,Ecevit Yilmaz,Thomas Laurell,Simon Ekström
Journal of Chromatography A. 2017; 1519: 1
[Pubmed] | [DOI]
5 A simple liquid extraction protocol for overcoming the ion suppression of triacylglycerols by phospholipids in liquid chromatography mass spectrometry studies
Pedro Araujo,Ephrem Tilahun,Joar Fjørtoft Breivik,Bashir M. Abdulkader,Livar Frøyland,Yingxu Zeng
Talanta. 2016; 148: 463
[Pubmed] | [DOI]
6 Novel advances in shotgun lipidomics for biology and medicine
Miao Wang,Chunyan Wang,Rowland H. Han,Xianlin Han
Progress in Lipid Research. 2016; 61: 83
[Pubmed] | [DOI]
7 Matrix removal in state of the art sample preparation methods for serum by charged aerosol detection and metabolomics-based LC-MS
Denise Schimek,Kevin A. Francesconi,Anton Mautner,Gunnar Libiseller,Reingard Raml,Christoph Magnes
Analytica Chimica Acta. 2016;
[Pubmed] | [DOI]
8 The impact of phospholipids and phospholipid removal on bioanalytical method performance
Jennifer Carmical,Stacy Brown
Biomedical Chromatography. 2016; 30(5): 710
[Pubmed] | [DOI]
9 Liquid chromatography mass spectrometry determination of perfluoroalkyl acids in environmental solid extracts after phospholipid removal and on-line turbulent flow chromatography purification
M. Mazzoni,S. Polesello,M. Rusconi,S. Valsecchi
Journal of Chromatography A. 2016;
[Pubmed] | [DOI]
10 The application of phospholipid removal columns and ultra-high performance liquid chromatography - tandem quadrupole mass spectrometry for quantification of multi-class antibiotics in aquaculture samples
Ingars Reinholds,Iveta Pugajeva,Ingus Perkons,Vadims Bartkevics
Journal of Pharmaceutical and Biomedical Analysis. 2016;
[Pubmed] | [DOI]
11 Templated polymers enable selective capture and release of lysophosphatidic acid in human plasma via optimization of non-covalent binding to functional monomers
Jialu Wang,Martha Sibrian-Vazquez,Jorge O. Escobedo,Lei Wang,Yu-Hsuan Chu,Richard G. Moore,Robert M. Strongin
The Analyst. 2015; 140(22): 7572
[Pubmed] | [DOI]
12 Comprehensive and Quantitative Analysis of Lysophospholipid Molecular Species Present in Obese Mouse Liver by Shotgun Lipidomics
Chunyan Wang,Miao Wang,Xianlin Han
Analytical Chemistry. 2015; 87(9): 4879
[Pubmed] | [DOI]
13 Development of an Automated Column Solid-Phase Extraction Cleanup of QuEChERS Extracts, Using a Zirconia-Based Sorbent, for Pesticide Residue Analyses by LC-MS/MS
Bruce D. Morris,Richard B. Schriner
Journal of Agricultural and Food Chemistry. 2015; 63(21): 5107
[Pubmed] | [DOI]
14 Mass spectrometric imaging of brain tissue by time-of-flight secondary ion mass spectrometry - How do polyatomic primary beams C60+, Ar2000+, water-doped Ar2000+and (H2O)6000+compare?
Irma Berrueta Razo,Sadia Sheraz,Alex Henderson,Nicholas P. Lockyer,John C. Vickerman
Rapid Communications in Mass Spectrometry. 2015; 29(20): 1851
[Pubmed] | [DOI]
15 Reverse-Polynomial Dilution Calibration Methodology Extends Lower Limit of Quantification and Reduces Relative Residual Error in Targeted Peptide Measurements in Blood Plasma
Yunki Y. Yau,Xizi Duo,Rupert W.L. Leong,Valerie C. Wasinger
Molecular & Cellular Proteomics. 2015; 14(2): 441
[Pubmed] | [DOI]
16 Analytical quality assurance in veterinary drug residue analysis methods: matrix effects determination and monitoring for sulfonamides analysis
Rodrigo Barcellos Hoff,Gabriel Rübensam,Louise Jank,Fabiano Barreto,Maria do Carmo Ruaro Peralba,Tânia Mara Pizzolato,M. Silvia Díaz-Cruz,Damià Barceló
Talanta. 2014;
[Pubmed] | [DOI]
17 Bioanalytical chromatographic method validation according to current regulations, with a special focus on the non-well defined parameters limit of quantification, robustness and matrix effect
Oskar González,María Encarnación Blanco,Gorka Iriarte,Luis Bartolomé,Miren Itxaso Maguregui,Rosa Ma. Alonso
Journal of Chromatography A. 2014;
[Pubmed] | [DOI]
18 Recent advances in sample preparation techniques to overcome difficulties encountered during quantitative analysis of small molecules from biofluids using LC-MS/MS
Caroline Bylda,Roland Thiele,Uwe Kobold,Dietrich A. Volmer
The Analyst. 2014; 139(10): 2265
[Pubmed] | [DOI]
19 Phospholipid removal combined with a semi-automated 96-well SPE application for determination of budesonide in human plasma with LC-MS/MS
Karin Nilsson,Maria Andersson,Olof Beck
Journal of Chromatography B. 2014;
[Pubmed] | [DOI]
20 ASPECTS: an automation-assisted SPE method development system
Ming Li,Judy Chou,Kristopher W King,Liyu Yang
Bioanalysis. 2013; 5(13): 1661
[Pubmed] | [DOI]
21 ASPECTS: An automation-assisted SPE method development system
Li, M., Chou, J., King, K.W., Yang, L.
Bioanalysis. 2013; 5(13): 1661-1676
[Pubmed]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Matrix Effect
    How to Diagnose ...
    A Post-Column In...
    Problems/Difficu...
    Novel-Hybrid SPE...
    Method Developme...
    Summary and Conc...
   Acknowledgements
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3593    
    Printed127    
    Emailed0    
    PDF Downloaded105    
    Comments [Add]    
    Cited by others 21    

Recommend this journal