ORIGINAL RESEARCH ARTICLE
Year : 2016  |  Volume : 8  |  Issue : 5  |  Page : 100-104  

Cumulative effect of microwave sterilization on the physical properties of microwave polymerized and conventional heat-polymerized acrylic resin


1 Department of Prosthodontics, Vivekanandha Dental College for Women, Tiruchengode, Tamil Nadu, India
2 Department of Prosthodontics, Oxford Dental College, Bengaluru, Karnataka, India
3 Department of Prosthodontics, Rajah Muthaiah Dental College, Chidambaram, Tamil Nadu, India
4 Department of Periodontics, J.K.K. Nataraja Dental College, Komarapalayam, Tamil Nadu, India

Date of Submission06-Apr-2016
Date of Decision28-Apr-2016
Date of Acceptance06-May-2016
Date of Web Publication12-Oct-2016

Correspondence Address:
Dr. S Karthikeyan
Department of Prosthodontics, Vivekanandha Dental College for Women, Tiruchengode, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0975-7406.191935

Rights and Permissions
   Abstract 

Aims: To evaluate and compare the flexural strength and impact strength of conventional and microwave cured denture base resins before and after repeated sterilization using microwave energy to consider microwave curing as an alternative to the conventional method of sterilization. Materials and Methods: The conventional heat cure acrylic resin (DPI heat cure material) Group A and microwave-polymerized acrylic resin (Vipi Wave Acrylic resin) Group B were used to fabricate 100 acrylic resins samples using a standard metal die of (86 mm × 11 mm × 3 mm) dimensions. The criterion was flexural strength and impact strength testing which had Group A and Group B samples; 50 samples for flexural strength and 50 samples for impact strength measurement. For each criterion, five control samples were taken for Group A and Group B. The samples were stored in water before experimenting. The test samples were subject to four cycles of microwave sterilization; followed by flexural strength testing with a 3-point flexural test in universal testing machine (UNITEK 94100) and impact strength testing with impact testing machine (ENKAY Pr09/E1/16). Results: The physical properties had significant changes for conventionally cured denture base resins, whereas no changes found for microwave-cured resins after repeated sterilization cycles.

Keywords: Denture base resins, flexural strength, impact strength, microwave sterilization


How to cite this article:
Shafeeq S M, Karthikeyan S, Reddy SM, Karthigeyan S, Manikandan R, Thangavelu A. Cumulative effect of microwave sterilization on the physical properties of microwave polymerized and conventional heat-polymerized acrylic resin. J Pharm Bioall Sci 2016;8, Suppl S1:100-4

How to cite this URL:
Shafeeq S M, Karthikeyan S, Reddy SM, Karthigeyan S, Manikandan R, Thangavelu A. Cumulative effect of microwave sterilization on the physical properties of microwave polymerized and conventional heat-polymerized acrylic resin. J Pharm Bioall Sci [serial online] 2016 [cited 2021 Apr 16];8, Suppl S1:100-4. Available from: https://www.jpbsonline.org/text.asp?2016/8/5/100/191935

The practice of removable prosthodontics has increased the potential of cross-contamination between dental office and laboratory. Handling of the contaminated or potentially contaminated items may result in self-inoculation and transmission of microorganisms to laboratory personnel or other patient's dentures and prosthetic appliances.[1]

Methacrylate polymers are preferred material for the fabrication of removable dental prostheses. Adequate maintenance of removable prostheses is needed for denture wearers to have an esthetic, odor-free appliance, and good oral health. Various maintenance procedures by chemical disinfection may alter some material properties during cleaning and disinfection.

As a simple alternative to overcome the disadvantages of chemical disinfection on prosthesis, the use of microwave energy has been suggested. Microwave energy has currently been used as an effective method of sterilization, yet its effect on resin properties after repeated procedures is not clear. The purpose of this study was to evaluate and to compare the effect of repeated microwave sterilization on physical properties of microwave-polymerized and conventional heat-polymerized acrylic resins.


   Materials and Methods Top


The materials for the study were two acrylic denture base materials, conventional heat cure acrylic resin (DPI heat cure material) [Figure 1] and microwave-polymerized acrylic resin (Vipi Wave Acrylic resin) [Figure 2]. In this study, bar samples of dimensions of 86 mm × 11 mm × 3 mm were used which was adapted from the previous studies done by Zappini et al .[2] and Meng and Latta 2005[3] [Figure 3]. The die duplication material was type III dental stone and conventional brass flask and fiber reinforced flask (Vipi STG Flask) [Figure 4] was used for conventional heat cure resin and microwave cure resin, respectively. The resins were manipulated as per manufacturer's recommendation, packed in dough stage followed by trial closure, and excess flash removal. For conventional heat cure resin polymerization, the flask was immersed in an acrylizer and processed in the temperature of 74° for 9 h and terminal boiling for 30 min (Long curing cycle). Specimens were bench cooled for overnight before deflasking. For microwave polymerization, the FRP flask were kept in the turn table of the domestic microwave oven (LG - model Number 1927C, Max 800 W with the frequency of 2450 MHz) [Figure 5] with 90 W for 6 1/2 min and the flask was reversed to top down position and again for 6 1/2 min. Then, the power increased to 360 W for 3 min.[4] Specimens were bench cooled for overnight before deflasking.
Figure 1: Conventional heat cure acrylic resin

Click here to view
Figure 2: Microwave-polymerized acrylic resin

Click here to view
Figure 3: Metal dies used (86 mm × 11 mm × 3 mm)

Click here to view
Figure 4: Microwave flask and conventional flask with mold space created

Click here to view
Figure 5: Microwave oven (LG- 1927C)

Click here to view


A total of 100 samples were fabricated, finished, and polished [Figure 6]; 50 were conventional heat cured resin samples (Group A) and 50 were microwave-cured resin samples (Group B). The criterion was to test flexural strength and impact strength for the samples. Each criterion had 25 Group A and 25 Group B samples for testing. Prior to testing the specimens were stored in water at 37°C for 24 h. This was done because water sorption causes water mobility among resin molecules, and the stress induced during curing can be relaxed resulting in intermolecular relaxation.[5]
Figure 6: Finished and polished acrylic samples

Click here to view


Microwave sterilization

Fifty samples were for testing flexural strength and fifty samples for impact strength. Sampling was systemized for each testing criterion as below.

In Group A and Group B individually, 5 samples were the control which were not sterilized (T0), and 5 samples were microwave irradiated one time for 3 min at 650 W (T1) after 7 days from storage in water, another 5 samples were microwave irradiated second time after 15 days following the same protocol (T2), another 5 samples were microwave irradiated third time after 21 days following the same protocol (T3), and another 5 samples were microwave irradiated fourth time after 30 days following the same protocol (T4). A total of 20 control samples (T0) without microwave sterilization and 80 samples (T1, T2, T3, and T4) with periodical microwave sterilization were segregated for testing. The tested disinfection protocol followed the study by Mima et al .,[4] who showed that microwave irradiation at 650 W for 3 min was effective to sterilize hard chairside reline resins. The same settings were adopted for this study.

Flexural strength testing

Five bars from each group were tested using a 3-point flexural test [Figure 7] (ISO 1567; specification for denture base polymers) for control group and sample groups after each cycle of sterilization. Specimens were placed in a test rig with circular supports 50 mm apart. The plunger and supports were 3 mm in diameter and 10 mm long. A force was applied using a universal testing machine (UNITEK 94100) and a 500-lbs load cell at a crosshead speed of 5 mm/min. The fracture force was recorded in newtons (N), and the flexural strength (F) was calculated per the formula to yield MPa units (Flexural strength) F = 3 PL/2 bd 2. WhereP is maximum load (fracture load), L is span length (86 mm), b is specimen width (11 mm), d is specimen thickness (3 mm).
Figure 7: Three-point bending test with universal testing machine

Click here to view


Impact strength testing

Five bars from each group of samples were tested using an impact testing machine for control group and sample groups after each cycle of sterilization. Specimens were placed in a test slot with a small portion projecting out. When the hammer of the impact testing machine (ENKAY Pr09/E1/16) [Figure 8] hits the specimen, the reading at which the specimen breaks gives the impact strength.
Figure 8: Impact testing machine (ENKAY PR 09/E1/16)

Click here to view


Readings for control samples and samples after each cycle of sterilization were recorded and noted separately for flexural strength and impact strength.


   Results Top


Statistical analyses were completed using Student's t -test to find out the significant difference between the two groups. If P < 0.05 then it is statistically significant. If the P > 0.05 then it is not statistically significant [Table 1] and [Table 2]. The physical properties had significant changes for conventionally cured denture base resins, whereas no changes found for microwave-cured resins after repeated sterilization cycles.
Table 1: The t -test results for impact and flexural strength measurements of the control samples (T0) versus study group samples (T1, T2, T3, and T4) of conventional heat cured acrylic resin

Click here to view
Table 2: The t -test results for impact and flexural strength measurements of the control samples (T0) versus study group samples (T1, T2, T3, and T4) microwave-cured acrylic resin

Click here to view



   Discussion Top


Acrylic resins have been used for denture fabrication for over 60 years. Oral problems related to poor hygiene of dentures support the need to establish a disinfection protocol that is effective, clinically viable, inexpensive, and easy to comply with.[6]

Chemical disinfection of dentures is commonly achieved by soaking in alkaline glutaraldehyde, sodium hypochlorite, aqueous formaldehyde, or enzymatic solutions. These are still controversial because they may alter some material properties and may induce resistance to candida albicans . Some alcohol-based disinfectants reduce the flexural strength of non-crosslinked denture base acrylic resins. Surface alteration and roughness may occur by continuous use of some disinfection methods, for example, staining (by soaking in chlorhexidine) or bleaching (by soaking in sodium hypochlorite).

As a simple alternative to overcome the disadvantages of chemical disinfection on prosthesis, the use of microwave energy has been suggested. It is considered an easily accessible, simple to perform, and effective low-cost method, yet its effect on resin properties after repeated procedures is not clear.

Webb et al . in 1998 did a study to test the in vitro efficacy of two methods of denture sterilization (microwave irradiation and sodium hypochlorite soak). The results of his study indicated that microwaving may be a more effective method of denture sterilization than denture soaking in sodium hypochlorite.[7] There have been a number of studies carried out by other investigators in relation to the sterilization of microorganisms by microwave irradiation (Rohrer et al ., 1985)[8],[9],[10] but (Rohrer and Bulard, 1985) was concerned with the sterilization of microorganisms on denture surfaces and its effects on the dentures.

In this study, both conventional and microwave resins were checked for flexural strength and impact strength before sterilization and after repeated four cycles of microwave sterilization. It was observed in conventional resin there was decrease in flexural strength after repeated cycles of sterilization and there was statistically significant increase in impact strength after repeated cycles of microwave sterilization. This might be due to the release of internal stresses introduced during processing and curing.[11] Whereas in microwave resin, there was statistically neither change in flexural strength or impact strength, possibly because there may not be much stress induced while microwave processing and curing, unlike conventional resin. Microwave heating is independent of thermal conductivity, and the temperature rises fast in the center of the resin bulk. Therefore, application of rapid heat may be used without the development of a high exothermic temperature.[12]

The clinical implication of this study is that microwave sterilization is an alternate method of denture sterilization for microwave-cured acrylic resins which can be carried out without any significant alteration in the physical properties of the denture base resins even for long-term use.

The limitations of these studies are that it has utilized only a specific shape of the specimen and complex shapes such as complete dentures are not used for this study. Further studies with such complex shapes are necessary to confirm the results of this study. Further cycles of sterilization should be checked for dimensional change, flexural properties, impact strength, to confirm the results of this study.


   Conclusion Top


Within the limitations of this study, it was concluded that microwave sterilization is an alternative method of routine sterilization for microwave cured complete dentures, without any change in the physical properties of the denture base material and it is not a suitable method for sterilization of conventionally cured complete dentures.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Pavan S, Arioli Filho JN, Dos Santos PH, Mollo Fde A Jr. Effect of microwave treatments on dimensional accuracy of maxillary acrylic resin denture base. Braz Dent J 2005;16:119-23.  Back to cited text no. 1
[PUBMED]    
2.
Zappini G, Kammann A, Wachter W. Comparision of fracture tests of denture base materials. J Prosthet Dent 2003;90:578-85.  Back to cited text no. 2
[PUBMED]    
3.
Meng TR Jr., Latta MA. Physical properties of four acrylic denture base resins. J Contemp Dent Pract 2005;6:93-100.  Back to cited text no. 3
    
4.
Mima EG, Pavarina AC, Neppelenbroek KH, Vergani CE, Spolidorio DM, Machado AL. Effect of different exposure times on microwave irradiation on the disinfection of a hard chairside reline resin. J Prosthodont 2008;17:312-7.  Back to cited text no. 4
[PUBMED]    
5.
Sartori EA, Schmidt CB, Mota EG, Hirakata LM, Shinkai RS. Cumulative effect of disinfection procedures on microhardness and tridimensional stability of a poly(methyl methacrylate) denture base resin. J Biomed Mater Res B Appl Biomater 2008;86:360-4.  Back to cited text no. 5
[PUBMED]    
6.
Sartori EA, Schmidt CB, Walber LF, Shinkai RS. Effect of microwave disinfection on denture base adaptation and resin surface roughness. Braz Dent J 2006;17:195-200.  Back to cited text no. 6
[PUBMED]    
7.
Webb BC, Thomas CJ, Harty DW, Willcox MD. Effectiveness of two methods of denture sterilization. J Oral Rehabil 1998;25:416-23.  Back to cited text no. 7
[PUBMED]    
8.
Young SK, Graves DC, Rohrer MD, Bulard RA. Microwave sterilization of nitrous oxide nasal hoods contaminated with virus. Oral Surg Oral Med Oral Pathol 1985;60:581-5.  Back to cited text no. 8
[PUBMED]    
9.
Rohrer MD, Bulard RA. Microwave sterilization. J Am Dent Assoc 1985;110:194-8.  Back to cited text no. 9
[PUBMED]    
10.
Dixon DL, Breeding LC, Faler TA. Microwave disinfection of denture base materials colonized with Candida albicans . J Prosthet Dent 1999;81:207-14.  Back to cited text no. 10
[PUBMED]    
11.
Ganzarolli SM, de Mello JA, Shinkai RS, Del Bel Cury AA. Internal adaptation and some physical properties of methacrylate-based denture base resins polymerized by different techniques. J Biomed Mater Res B Appl Biomater 2007;82:169-73.  Back to cited text no. 11
[PUBMED]    
12.
Levin B, Sanders JL, Reitz PV. The use of microwave energy for processing acrylic resins. J Prosthet Dent 1989;61:381-3.  Back to cited text no. 12
[PUBMED]    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
    Materials and Me...
   Results
   Discussion
   Conclusion
    References